```markdown # Goal/Experiment: To acquire photosynthetic efficiency (Fv/Fm) and quantum yield of photosystem II (Y(II)) of Microcystis aeruginosa cultures using far-red acclimation with the PHYO-PAM-II Compact Version. # Application of PHYTO-PAM-II (Compact Version) For Running Rapid Light Curves on Cyanobacterial Samples **Forked from**: [Application of PHYTO-PAM-II (Compact Version) on Aureococcus anophagefferens cultures for photosynthetic efficiency and quantum yield of PSII](https://dx.doi.org/10.17504/protocols.io.eq2ly7jqelx9/v1) ### Citation: Gwen Stark, Emily E. Chase, Steven W. Wilhelm 2022. Application of PHYTO-PAM-II (Compact Version) For Running Rapid light curves on Cyanobacterial samples. protocols.io https://dx.doi.org/10.17504/protocols.io.eq2ly7jqelx9/v1 ### Contributors: Gwen Stark1, Emily E. Chase2, Steven W. Wilhelm1 1. The University of Tennessee, Knoxville 2. University of Tennessee, Knoxville --- ## Abstract A protocol to acquire photosynthetic efficiency (Fv/Fm) and quantum yield of photosystem II (Y(II)) of Microcystis aeruginosa cultures using far-red acclimation. ## Keywords - Photosynthetic efficiency - Quantum yield - Fv/Fm - HAB - Chlorophyll fluorescence - Cyanobacteria ## License This is an open access protocol distributed under the terms of the [Creative Commons Attribution License](https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. --- ## Guidelines The collection of quantum yield and photosynthetic efficiency is highly sensitive to modifications in sampling protocol. ## Materials - PHYTO-PAM-II Compact Version and components - Laptop with a USB port, and PhytoWin_3 installed - Measurement cuvettes - At least 3 mL of culture material for each sample - 70% Ethanol - KimWipes - Necessary materials and setup for both dark adapting cultures, and a no light or low light environment for sampling ## Before Starting Familiarize yourself with the PHYTO-PAM-II equipment, manufacturer’s provided manual, and the basics of chlorophyll fluorescence parameters for full use of data collected and accuracy of results. --- ## Equipment Preparation 1. **Setting Up the PHYTO-PAM-II Compact Version:** - Toggle the power and connect to a laptop via USB. - Open the PhytoWin_3 program and select "Phyto Compact Unit." - Ensure the program is in "MEASURE" mode, and that the "ML" light is green (indicating it is ready). - Ensure that "AL" and "FR" lights are not selected. ![Equipment Setup](image1.png) 2. **Adjusting Settings:** - Modify Phyto-Pam settings based on the species being tested. - Use the "View Pulse" check box to observe the saturation pulse kinetics, ensuring a distinct plateau is seen. --- ## Sample Preparation 3. **Sample Transfer:** - Transfer 3 mL of culture to a quartz cuvette and minimize exposure time. - Place the cuvette in the optical port and cover with the darkening hood. 4. **Auto-Adjust Gain:** - Perform a zero-offset function using Zoff-determination to blank the PHYTO-PAM-II before runs. 5. **Far-Red Acclimation:** - Apply far-red light acclimation to induce photosystem I and oxidize the PQ pool. - Dark acclimation is not recommended for cyanobacteria. > **Note:** Evaluate the necessity between no acclimation and weak far-red acclimation for consistent results. --- ## Measurement Acquisition 6. **Experimental Measurements:** - Measurements can now be collected. 7. **Running Far-Red Light Acclimation:** - Transfer 3 mL of culture into the cuvette, auto-adjust gain, and run acclimation. - Turn on the "ML" button and wait for the light to turn green. 8. **Taking Measurements:** - Choose Fv/Fm measurements or run a rapid light curve. - Adjust PAR and exposure times appropriately. - Aim for the ETR curve to hit a maximum peak and level off. 9. **Recording Data:** - Data will be recorded automatically. - Use 70% ethanol to clean the cuvette between different treatments. 10. **Data Storage:** - Save and export data as a .csv file. - Unplug the equipment and charge fully for long-term storage. > **Recommendation:** Regularly charge the equipment if stored unused for extended periods. --- ### endofoutput ```