Chapter
stringclasses
18 values
sentence_range
stringlengths
3
9
Text
stringlengths
7
7.34k
1
3990-3993
[4 31(a)] is very similar to an expression obtained earlier for the electric field of a dipole The similarity may be seen if we substitute, 0 1/0 µ ε → Rationalised 2023-24 129 Moving Charges and Magnetism e m→ p (electrostatic dipole) B→ E (electrostatic field) We then obtain, 3 0 2 4 e εx = π p E which is precisely the field for an electric dipole at a point on its axis considered in Chapter 1, Section 1
1
3991-3994
31(a)] is very similar to an expression obtained earlier for the electric field of a dipole The similarity may be seen if we substitute, 0 1/0 µ ε → Rationalised 2023-24 129 Moving Charges and Magnetism e m→ p (electrostatic dipole) B→ E (electrostatic field) We then obtain, 3 0 2 4 e εx = π p E which is precisely the field for an electric dipole at a point on its axis considered in Chapter 1, Section 1 10 [Eq
1
3992-3995
The similarity may be seen if we substitute, 0 1/0 µ ε → Rationalised 2023-24 129 Moving Charges and Magnetism e m→ p (electrostatic dipole) B→ E (electrostatic field) We then obtain, 3 0 2 4 e εx = π p E which is precisely the field for an electric dipole at a point on its axis considered in Chapter 1, Section 1 10 [Eq (1
1
3993-3996
considered in Chapter 1, Section 1 10 [Eq (1 20)]
1
3994-3997
10 [Eq (1 20)] It can be shown that the above analogy can be carried further
1
3995-3998
(1 20)] It can be shown that the above analogy can be carried further We had found in Chapter 1 that the electric field on the perpendicular bisector of the dipole is given by [See Eq
1
3996-3999
20)] It can be shown that the above analogy can be carried further We had found in Chapter 1 that the electric field on the perpendicular bisector of the dipole is given by [See Eq (1
1
3997-4000
It can be shown that the above analogy can be carried further We had found in Chapter 1 that the electric field on the perpendicular bisector of the dipole is given by [See Eq (1 21)], E ≃ pe x 4 0 3 πε where x is the distance from the dipole
1
3998-4001
We had found in Chapter 1 that the electric field on the perpendicular bisector of the dipole is given by [See Eq (1 21)], E ≃ pe x 4 0 3 πε where x is the distance from the dipole If we replace p à m and 0 1/0 µ ε → in the above expression, we obtain the result for B for a point in the plane of the loop at a distance x from the centre
1
3999-4002
(1 21)], E ≃ pe x 4 0 3 πε where x is the distance from the dipole If we replace p à m and 0 1/0 µ ε → in the above expression, we obtain the result for B for a point in the plane of the loop at a distance x from the centre For x >>R, B m ≃ µ0 4π x3 x R ; >> [4
1
4000-4003
21)], E ≃ pe x 4 0 3 πε where x is the distance from the dipole If we replace p à m and 0 1/0 µ ε → in the above expression, we obtain the result for B for a point in the plane of the loop at a distance x from the centre For x >>R, B m ≃ µ0 4π x3 x R ; >> [4 31(b)] The results given by Eqs
1
4001-4004
If we replace p à m and 0 1/0 µ ε → in the above expression, we obtain the result for B for a point in the plane of the loop at a distance x from the centre For x >>R, B m ≃ µ0 4π x3 x R ; >> [4 31(b)] The results given by Eqs [4
1
4002-4005
For x >>R, B m ≃ µ0 4π x3 x R ; >> [4 31(b)] The results given by Eqs [4 31(a)] and [4
1
4003-4006
31(b)] The results given by Eqs [4 31(a)] and [4 31(b)] become exact for a point magnetic dipole
1
4004-4007
[4 31(a)] and [4 31(b)] become exact for a point magnetic dipole The results obtained above can be shown to apply to any planar loop: a planar current loop is equivalent to a magnetic dipole of dipole moment m = I A, which is the analogue of electric dipole moment p
1
4005-4008
31(a)] and [4 31(b)] become exact for a point magnetic dipole The results obtained above can be shown to apply to any planar loop: a planar current loop is equivalent to a magnetic dipole of dipole moment m = I A, which is the analogue of electric dipole moment p Note, however, a fundamental difference: an electric dipole is built up of two elementary units — the charges (or electric monopoles)
1
4006-4009
31(b)] become exact for a point magnetic dipole The results obtained above can be shown to apply to any planar loop: a planar current loop is equivalent to a magnetic dipole of dipole moment m = I A, which is the analogue of electric dipole moment p Note, however, a fundamental difference: an electric dipole is built up of two elementary units — the charges (or electric monopoles) In magnetism, a magnetic dipole (or a current loop) is the most elementary element
1
4007-4010
The results obtained above can be shown to apply to any planar loop: a planar current loop is equivalent to a magnetic dipole of dipole moment m = I A, which is the analogue of electric dipole moment p Note, however, a fundamental difference: an electric dipole is built up of two elementary units — the charges (or electric monopoles) In magnetism, a magnetic dipole (or a current loop) is the most elementary element The equivalent of electric charges, i
1
4008-4011
Note, however, a fundamental difference: an electric dipole is built up of two elementary units — the charges (or electric monopoles) In magnetism, a magnetic dipole (or a current loop) is the most elementary element The equivalent of electric charges, i e
1
4009-4012
In magnetism, a magnetic dipole (or a current loop) is the most elementary element The equivalent of electric charges, i e , magnetic monopoles, are not known to exist
1
4010-4013
The equivalent of electric charges, i e , magnetic monopoles, are not known to exist We have shown that a current loop (i) produces a magnetic field and behaves like a magnetic dipole at large distances, and (ii) is subject to torque like a magnetic needle
1
4011-4014
e , magnetic monopoles, are not known to exist We have shown that a current loop (i) produces a magnetic field and behaves like a magnetic dipole at large distances, and (ii) is subject to torque like a magnetic needle This led Ampere to suggest that all magnetism is due to circulating currents
1
4012-4015
, magnetic monopoles, are not known to exist We have shown that a current loop (i) produces a magnetic field and behaves like a magnetic dipole at large distances, and (ii) is subject to torque like a magnetic needle This led Ampere to suggest that all magnetism is due to circulating currents This seems to be partly true and no magnetic monopoles have been seen so far
1
4013-4016
We have shown that a current loop (i) produces a magnetic field and behaves like a magnetic dipole at large distances, and (ii) is subject to torque like a magnetic needle This led Ampere to suggest that all magnetism is due to circulating currents This seems to be partly true and no magnetic monopoles have been seen so far However, elementary particles such as an electron or a proton also carry an intrinsic magnetic moment, not accounted by circulating currents
1
4014-4017
This led Ampere to suggest that all magnetism is due to circulating currents This seems to be partly true and no magnetic monopoles have been seen so far However, elementary particles such as an electron or a proton also carry an intrinsic magnetic moment, not accounted by circulating currents 4
1
4015-4018
This seems to be partly true and no magnetic monopoles have been seen so far However, elementary particles such as an electron or a proton also carry an intrinsic magnetic moment, not accounted by circulating currents 4 10 THE MOVING COIL GALVANOMETER Currents and voltages in circuits have been discussed extensively in Chapters 3
1
4016-4019
However, elementary particles such as an electron or a proton also carry an intrinsic magnetic moment, not accounted by circulating currents 4 10 THE MOVING COIL GALVANOMETER Currents and voltages in circuits have been discussed extensively in Chapters 3 But how do we measure them
1
4017-4020
4 10 THE MOVING COIL GALVANOMETER Currents and voltages in circuits have been discussed extensively in Chapters 3 But how do we measure them How do we claim that current in a circuit is 1
1
4018-4021
10 THE MOVING COIL GALVANOMETER Currents and voltages in circuits have been discussed extensively in Chapters 3 But how do we measure them How do we claim that current in a circuit is 1 5 A or the voltage drop across a resistor is 1
1
4019-4022
But how do we measure them How do we claim that current in a circuit is 1 5 A or the voltage drop across a resistor is 1 2 V
1
4020-4023
How do we claim that current in a circuit is 1 5 A or the voltage drop across a resistor is 1 2 V Figure 4
1
4021-4024
5 A or the voltage drop across a resistor is 1 2 V Figure 4 20 exhibits a very useful instrument for this purpose: the moving coil galvanometer (MCG)
1
4022-4025
2 V Figure 4 20 exhibits a very useful instrument for this purpose: the moving coil galvanometer (MCG) It is a device whose principle can be understood on the basis of our discussion in Section 4
1
4023-4026
Figure 4 20 exhibits a very useful instrument for this purpose: the moving coil galvanometer (MCG) It is a device whose principle can be understood on the basis of our discussion in Section 4 10
1
4024-4027
20 exhibits a very useful instrument for this purpose: the moving coil galvanometer (MCG) It is a device whose principle can be understood on the basis of our discussion in Section 4 10 The galvanometer consists of a coil, with many turns, free to rotate about a fixed axis (Fig
1
4025-4028
It is a device whose principle can be understood on the basis of our discussion in Section 4 10 The galvanometer consists of a coil, with many turns, free to rotate about a fixed axis (Fig 4
1
4026-4029
10 The galvanometer consists of a coil, with many turns, free to rotate about a fixed axis (Fig 4 20), in a uniform radial magnetic field
1
4027-4030
The galvanometer consists of a coil, with many turns, free to rotate about a fixed axis (Fig 4 20), in a uniform radial magnetic field There is a cylindrical soft iron core which not only makes the field radial but also increases the strength of the magnetic field
1
4028-4031
4 20), in a uniform radial magnetic field There is a cylindrical soft iron core which not only makes the field radial but also increases the strength of the magnetic field When a current flows through the coil, a torque acts on it
1
4029-4032
20), in a uniform radial magnetic field There is a cylindrical soft iron core which not only makes the field radial but also increases the strength of the magnetic field When a current flows through the coil, a torque acts on it This torque is given by Eq
1
4030-4033
There is a cylindrical soft iron core which not only makes the field radial but also increases the strength of the magnetic field When a current flows through the coil, a torque acts on it This torque is given by Eq (4
1
4031-4034
When a current flows through the coil, a torque acts on it This torque is given by Eq (4 26) to be t = NI AB Rationalised 2023-24 Physics 130 where the symbols have their usual meaning
1
4032-4035
This torque is given by Eq (4 26) to be t = NI AB Rationalised 2023-24 Physics 130 where the symbols have their usual meaning Since the field is radial by design, we have taken sin q = 1 in the above expression for the torque
1
4033-4036
(4 26) to be t = NI AB Rationalised 2023-24 Physics 130 where the symbols have their usual meaning Since the field is radial by design, we have taken sin q = 1 in the above expression for the torque The magnetic torque NIAB tends to rotate the coil
1
4034-4037
26) to be t = NI AB Rationalised 2023-24 Physics 130 where the symbols have their usual meaning Since the field is radial by design, we have taken sin q = 1 in the above expression for the torque The magnetic torque NIAB tends to rotate the coil A spring Sp provides a counter torque kf that balances the magnetic torque NIAB; resulting in a steady angular deflection f
1
4035-4038
Since the field is radial by design, we have taken sin q = 1 in the above expression for the torque The magnetic torque NIAB tends to rotate the coil A spring Sp provides a counter torque kf that balances the magnetic torque NIAB; resulting in a steady angular deflection f In equilibrium kf = NI AB where k is the torsional constant of the spring; i
1
4036-4039
The magnetic torque NIAB tends to rotate the coil A spring Sp provides a counter torque kf that balances the magnetic torque NIAB; resulting in a steady angular deflection f In equilibrium kf = NI AB where k is the torsional constant of the spring; i e
1
4037-4040
A spring Sp provides a counter torque kf that balances the magnetic torque NIAB; resulting in a steady angular deflection f In equilibrium kf = NI AB where k is the torsional constant of the spring; i e the restoring torque per unit twist
1
4038-4041
In equilibrium kf = NI AB where k is the torsional constant of the spring; i e the restoring torque per unit twist The deflection f is indicated on the scale by a pointer attached to the spring
1
4039-4042
e the restoring torque per unit twist The deflection f is indicated on the scale by a pointer attached to the spring We have φ =   NAB k I (4
1
4040-4043
the restoring torque per unit twist The deflection f is indicated on the scale by a pointer attached to the spring We have φ =   NAB k I (4 38) The quantity in brackets is a constant for a given galvanometer
1
4041-4044
The deflection f is indicated on the scale by a pointer attached to the spring We have φ =   NAB k I (4 38) The quantity in brackets is a constant for a given galvanometer The galvanometer can be used in a number of ways
1
4042-4045
We have φ =   NAB k I (4 38) The quantity in brackets is a constant for a given galvanometer The galvanometer can be used in a number of ways It can be used as a detector to check if a current is flowing in the circuit
1
4043-4046
38) The quantity in brackets is a constant for a given galvanometer The galvanometer can be used in a number of ways It can be used as a detector to check if a current is flowing in the circuit We have come across this usage in the Wheatstone’s bridge arrangement
1
4044-4047
The galvanometer can be used in a number of ways It can be used as a detector to check if a current is flowing in the circuit We have come across this usage in the Wheatstone’s bridge arrangement In this usage the neutral position of the pointer (when no current is flowing through the galvanometer) is in the middle of the scale and not at the left end as shown in Fig
1
4045-4048
It can be used as a detector to check if a current is flowing in the circuit We have come across this usage in the Wheatstone’s bridge arrangement In this usage the neutral position of the pointer (when no current is flowing through the galvanometer) is in the middle of the scale and not at the left end as shown in Fig 4
1
4046-4049
We have come across this usage in the Wheatstone’s bridge arrangement In this usage the neutral position of the pointer (when no current is flowing through the galvanometer) is in the middle of the scale and not at the left end as shown in Fig 4 20
1
4047-4050
In this usage the neutral position of the pointer (when no current is flowing through the galvanometer) is in the middle of the scale and not at the left end as shown in Fig 4 20 Depending on the direction of the current, the pointer’s deflection is either to the right or the left
1
4048-4051
4 20 Depending on the direction of the current, the pointer’s deflection is either to the right or the left The galvanometer cannot as such be used as an ammeter to measure the value of the current in a given circuit
1
4049-4052
20 Depending on the direction of the current, the pointer’s deflection is either to the right or the left The galvanometer cannot as such be used as an ammeter to measure the value of the current in a given circuit This is for two reasons: (i) Galvanometer is a very sensitive device, it gives a full- scale deflection for a current of the order of mA
1
4050-4053
Depending on the direction of the current, the pointer’s deflection is either to the right or the left The galvanometer cannot as such be used as an ammeter to measure the value of the current in a given circuit This is for two reasons: (i) Galvanometer is a very sensitive device, it gives a full- scale deflection for a current of the order of mA (ii) For measuring currents, the galvanometer has to be connected in series, and as it has a large resistance, this will change the value of the current in the circuit
1
4051-4054
The galvanometer cannot as such be used as an ammeter to measure the value of the current in a given circuit This is for two reasons: (i) Galvanometer is a very sensitive device, it gives a full- scale deflection for a current of the order of mA (ii) For measuring currents, the galvanometer has to be connected in series, and as it has a large resistance, this will change the value of the current in the circuit To overcome these difficulties, one attaches a small resistance rs, called shunt resistance, in parallel with the galvanometer coil; so that most of the current passes through the shunt
1
4052-4055
This is for two reasons: (i) Galvanometer is a very sensitive device, it gives a full- scale deflection for a current of the order of mA (ii) For measuring currents, the galvanometer has to be connected in series, and as it has a large resistance, this will change the value of the current in the circuit To overcome these difficulties, one attaches a small resistance rs, called shunt resistance, in parallel with the galvanometer coil; so that most of the current passes through the shunt The resistance of this arrangement is, RG rs / (RG + rs) ≃ rs if RG >> rs If rs has small value, in relation to the resistance of the rest of the circuit Rc, the effect of introducing the measuring instrument is also small and negligible
1
4053-4056
(ii) For measuring currents, the galvanometer has to be connected in series, and as it has a large resistance, this will change the value of the current in the circuit To overcome these difficulties, one attaches a small resistance rs, called shunt resistance, in parallel with the galvanometer coil; so that most of the current passes through the shunt The resistance of this arrangement is, RG rs / (RG + rs) ≃ rs if RG >> rs If rs has small value, in relation to the resistance of the rest of the circuit Rc, the effect of introducing the measuring instrument is also small and negligible This arrangement is schematically shown in Fig
1
4054-4057
To overcome these difficulties, one attaches a small resistance rs, called shunt resistance, in parallel with the galvanometer coil; so that most of the current passes through the shunt The resistance of this arrangement is, RG rs / (RG + rs) ≃ rs if RG >> rs If rs has small value, in relation to the resistance of the rest of the circuit Rc, the effect of introducing the measuring instrument is also small and negligible This arrangement is schematically shown in Fig 4
1
4055-4058
The resistance of this arrangement is, RG rs / (RG + rs) ≃ rs if RG >> rs If rs has small value, in relation to the resistance of the rest of the circuit Rc, the effect of introducing the measuring instrument is also small and negligible This arrangement is schematically shown in Fig 4 21
1
4056-4059
This arrangement is schematically shown in Fig 4 21 The scale of this ammeter is calibrated and then graduated to read off the current value with ease
1
4057-4060
4 21 The scale of this ammeter is calibrated and then graduated to read off the current value with ease We define the current sensitivity of the galvanometer as the deflection per unit current
1
4058-4061
21 The scale of this ammeter is calibrated and then graduated to read off the current value with ease We define the current sensitivity of the galvanometer as the deflection per unit current From Eq
1
4059-4062
The scale of this ammeter is calibrated and then graduated to read off the current value with ease We define the current sensitivity of the galvanometer as the deflection per unit current From Eq (4
1
4060-4063
We define the current sensitivity of the galvanometer as the deflection per unit current From Eq (4 38) this current sensitivity is, NAB I k φ = (4
1
4061-4064
From Eq (4 38) this current sensitivity is, NAB I k φ = (4 39) A convenient way for the manufacturer to increase the sensitivity is to increase the number of turns N
1
4062-4065
(4 38) this current sensitivity is, NAB I k φ = (4 39) A convenient way for the manufacturer to increase the sensitivity is to increase the number of turns N We choose galvanometers having sensitivities of value, required by our experiment
1
4063-4066
38) this current sensitivity is, NAB I k φ = (4 39) A convenient way for the manufacturer to increase the sensitivity is to increase the number of turns N We choose galvanometers having sensitivities of value, required by our experiment FIGURE 4
1
4064-4067
39) A convenient way for the manufacturer to increase the sensitivity is to increase the number of turns N We choose galvanometers having sensitivities of value, required by our experiment FIGURE 4 20 The moving coil galvanometer
1
4065-4068
We choose galvanometers having sensitivities of value, required by our experiment FIGURE 4 20 The moving coil galvanometer Its elements are described in the text
1
4066-4069
FIGURE 4 20 The moving coil galvanometer Its elements are described in the text Depending on the requirement, this device can be used as a current detector or for measuring the value of the current (ammeter) or voltage (voltmeter)
1
4067-4070
20 The moving coil galvanometer Its elements are described in the text Depending on the requirement, this device can be used as a current detector or for measuring the value of the current (ammeter) or voltage (voltmeter) FIGURE 4
1
4068-4071
Its elements are described in the text Depending on the requirement, this device can be used as a current detector or for measuring the value of the current (ammeter) or voltage (voltmeter) FIGURE 4 21 Conversion of a galvanometer (G) to an ammeter by the introduction of a shunt resistance rs of very small value in parallel
1
4069-4072
Depending on the requirement, this device can be used as a current detector or for measuring the value of the current (ammeter) or voltage (voltmeter) FIGURE 4 21 Conversion of a galvanometer (G) to an ammeter by the introduction of a shunt resistance rs of very small value in parallel Rationalised 2023-24 131 Moving Charges and Magnetism The galvanometer can also be used as a voltmeter to measure the voltage across a given section of the circuit
1
4070-4073
FIGURE 4 21 Conversion of a galvanometer (G) to an ammeter by the introduction of a shunt resistance rs of very small value in parallel Rationalised 2023-24 131 Moving Charges and Magnetism The galvanometer can also be used as a voltmeter to measure the voltage across a given section of the circuit For this it must be connected in parallel with that section of the circuit
1
4071-4074
21 Conversion of a galvanometer (G) to an ammeter by the introduction of a shunt resistance rs of very small value in parallel Rationalised 2023-24 131 Moving Charges and Magnetism The galvanometer can also be used as a voltmeter to measure the voltage across a given section of the circuit For this it must be connected in parallel with that section of the circuit Further, it must draw a very small current, otherwise the voltage measurement will disturb the original set up by an amount which is very large
1
4072-4075
Rationalised 2023-24 131 Moving Charges and Magnetism The galvanometer can also be used as a voltmeter to measure the voltage across a given section of the circuit For this it must be connected in parallel with that section of the circuit Further, it must draw a very small current, otherwise the voltage measurement will disturb the original set up by an amount which is very large Usually we like to keep the disturbance due to the measuring device below one per cent
1
4073-4076
For this it must be connected in parallel with that section of the circuit Further, it must draw a very small current, otherwise the voltage measurement will disturb the original set up by an amount which is very large Usually we like to keep the disturbance due to the measuring device below one per cent To ensure this, a large resistance R is connected in series with the galvanometer
1
4074-4077
Further, it must draw a very small current, otherwise the voltage measurement will disturb the original set up by an amount which is very large Usually we like to keep the disturbance due to the measuring device below one per cent To ensure this, a large resistance R is connected in series with the galvanometer This arrangement is schematically depicted in Fig
1
4075-4078
Usually we like to keep the disturbance due to the measuring device below one per cent To ensure this, a large resistance R is connected in series with the galvanometer This arrangement is schematically depicted in Fig 4
1
4076-4079
To ensure this, a large resistance R is connected in series with the galvanometer This arrangement is schematically depicted in Fig 4 22
1
4077-4080
This arrangement is schematically depicted in Fig 4 22 Note that the resistance of the voltmeter is now, RG + R ≃ R : large The scale of the voltmeter is calibrated to read off the voltage value with ease
1
4078-4081
4 22 Note that the resistance of the voltmeter is now, RG + R ≃ R : large The scale of the voltmeter is calibrated to read off the voltage value with ease We define the voltage sensitivity as the deflection per unit voltage
1
4079-4082
22 Note that the resistance of the voltmeter is now, RG + R ≃ R : large The scale of the voltmeter is calibrated to read off the voltage value with ease We define the voltage sensitivity as the deflection per unit voltage From Eq
1
4080-4083
Note that the resistance of the voltmeter is now, RG + R ≃ R : large The scale of the voltmeter is calibrated to read off the voltage value with ease We define the voltage sensitivity as the deflection per unit voltage From Eq (4
1
4081-4084
We define the voltage sensitivity as the deflection per unit voltage From Eq (4 38), Vφ NAB k VI NAB k R =    =    1 (4
1
4082-4085
From Eq (4 38), Vφ NAB k VI NAB k R =    =    1 (4 40) An interesting point to note is that increasing the current sensitivity may not necessarily increase the voltage sensitivity
1
4083-4086
(4 38), Vφ NAB k VI NAB k R =    =    1 (4 40) An interesting point to note is that increasing the current sensitivity may not necessarily increase the voltage sensitivity Let us take Eq
1
4084-4087
38), Vφ NAB k VI NAB k R =    =    1 (4 40) An interesting point to note is that increasing the current sensitivity may not necessarily increase the voltage sensitivity Let us take Eq (4
1
4085-4088
40) An interesting point to note is that increasing the current sensitivity may not necessarily increase the voltage sensitivity Let us take Eq (4 39) which provides a measure of current sensitivity
1
4086-4089
Let us take Eq (4 39) which provides a measure of current sensitivity If N ® 2N, i
1
4087-4090
(4 39) which provides a measure of current sensitivity If N ® 2N, i e
1
4088-4091
39) which provides a measure of current sensitivity If N ® 2N, i e , we double the number of turns, then 2 I I φ φ → Thus, the current sensitivity doubles
1
4089-4092
If N ® 2N, i e , we double the number of turns, then 2 I I φ φ → Thus, the current sensitivity doubles However, the resistance of the galvanometer is also likely to double, since it is proportional to the length of the wire